| Validation
|
SteriSet containers are
CE-marked to demonstrate that they fulfill the general requirements of the
European Medical Device Directive. Part of the documentation which was
collected for this CE-mark are reports of tests performed according EN 868
part 8 “reusable sterilization containers for steam sterilization”:
|
| |
|
Annex E “Determination of
sterilization performance”
-
with full standardized metal
load
-
with fabric load
Annex F “Load dryness
tests”
-
with full standardized metal
load
-
with fabric load
|
| |
|
Historical evidence over the
past 30 years with different vacuum steam cycles acc. (old) German
standard DIN 58946 “steam sterilizers” do also propose a very high
level of confidence that sterility will be reached during a vacuum steam
cycle, provided the cycle is operated within his intended limits.
|
| |
|
In
the US, AMSCO American Sterilizer Company did comprehensive studies on SteriSet's compatibility with AMSCO's
US typical steam sterilization cycles for the 510k filing of SteriSet, including:
-
AMSCO
Medallion Series 270oF deep and pulsing vac
-
AMSCO
Eagle series 270oF prevac (pulsing) and express
-
AMSCOMATIC
Terminal 285 oF prevac (pulsing)
Details
are reported in "Eagle SteriSet Technical Report: Design and Qualification"
|
| |
These tests demonstrated that
SteriSet containers perform correct when sterilized with standard loads in a EN 285 resp. DIN resp. AMSCO steam
sterilizer. However, hospitals
are today in a situation, where more and more new instruments (new materials; holloware designs: example MIS
instruments) are used: these produce a general compatibility problem,
which can be expressed as follows:
- A
container manufacturer (a manufacturer of any packaging material / system) cannot evaluate all
possible load and cycle configurations which might ever be formed / combined out in the field.
He can only validate his own packaging / container with standardized test loads / instruments against
standardized - or well defined - sterilization cycles.
(Other packaging than containers, i.e. wrap or pouches, have the additional problem that you would need to
define the WAY OF APPLICATION : single wrapped, double wrapped ? How to form / tape the pack?)
- A
instrument manufacturer (specifically when we talk about ortho and neuro loaner instruments which come
in nicely organized - but very compact and hard to re-process organizing sets) might argument that he cannot
evaluate all possible PACKAGING and CYCLE configurations which might ever be formed / combined out in the
field, but could only validate his INSTRUMENT SET inside a RECOMMENDED packaging in a standardized - or well
defined - sterilization cycle.
- A
sterilizer manufacturer cannot evaluate all possible packaging / load configurations which might ever
be formed / combined out in the field, but can only validate his sterilizer / different cycles with
standardized loads (metal; fabric) in standardized packaging materials.
How
could any of the involved manufacturers - in view of a undefined potential of possible combinations
in "real life" - ever claim that his product would have been validated, that it would so be safe
to use in any combination in any hospital?
Claiming a "validation"
well knowing that at least one major parameter (i.e. the cycle) might be different in real
application would be less than half of the truth, might even lead to a false sense of security in the
hospitals!
This is the
background for the requirement of on site validation of all possible configuration: where
else than “on site” and how else than with the combination of packaging / load / cycle as used in daily
application could you validate a process?
|
| |
European
standard EN 554
gives here guidance on
how to formally validate a steam sterilization cycle.
Validation in that sense shall
produce documented evidence, that the tested packaging/load/cycle
configuration is capable to reproducably achieve sterile results.
Acc. EN554 that cannot be done by use of
chemical or biological indicators, but needs sensitive recording of the
temperatures inside a package during all stages of a cycle (thermocouples).
Hospitals all over Europe are
currently starting to validate their processes accordingly, but in most of
the cases do not have the equipment neither the expertise to perform as
required: consequently they use consultant companies or third party test
labs for that purpose.
|
| |
|
Note
In case of SteriSet you would
remove a handle on the short side (open the nuts inside – remove handle
screws) and insert the thermocouples through that hole. After the
validation you would re-assemble the handle.
|
| |
PCDs (Process Challenging Devices
To ease the
evaluation of “hollowware” (lumen devices) during steam sterilization there are currently ongoing standardization activities in Europe to develop “standardized hollowware test
packs”: PCDs (Process
Challenging Devices – described in EN 867-5 and the forthcoming international standard EN ISO 14937).
These would
define “worst case” situations for steam penetration into lumens (like i.e. a Bowie&Dick test pack defines
a worst case for air removal). PCDs shall be applied in every cycle / batch during routine operation: if at the
end of the cycle the indicator (biological or chemical) which is contained in the PCD signals “sterile” there
is a very high level of confidence that ALL PACKS which have been exposed in the same cycle are sterile.
Specifically in the absence of a thermometric validation the use of PCDs would drastically
reduce risks caused by possibly insufficient steam penetration into lumens. PCDs are also recommendable for
routine control of sterilization efficacy.
|